Medical device cybersecurity in hospital networks: legacy equipment and EU MDR requirements

Remote Admin Security Team·November 23, 2026·21 min read

Medical device cybersecurity in a hospital network is shared between two parties: the manufacturer answers for the design of the device and for approving patches, and the hospital answers for the network the device runs in, for access to it and for watching its traffic. You usually cannot update a CT scanner or a laboratory analyser with an old operating system yourself, but you can isolate it, limit its communication to what it needs and put remote servicing in order. Those three steps reduce the risk markedly without replacing any equipment.

This article covers the practical side of medical device cybersecurity: why clinical equipment runs on old systems, what belongs to the manufacturer and what to the provider, how to build an inventory and segment the network, and how to organise service access. The technical advice works in any country. The legal layer is EU-wide where it comes from the Medical Device Regulation (MDR) and GDPR, and we mark the passages that describe Polish law. In Poland the relevant law is the National Cybersecurity System Act (the KSC Act), which transposes NIS2; it was amended by the Act of 23 January 2026 (Journal of Laws 2026, item 252), in force since 3 April 2026. Legal status as of November 2026. This article is for information only and is not legal advice. If you operate in Poland and want to check how your equipment fits into your duties under the KSC Act, a good starting point is an NIS2/KSC audit whose scope can include the device network.

Why does medical equipment run on old operating systems?

Medical equipment runs on old operating systems because it lives much longer than the software it was built on, and every change to that software needs the manufacturer’s approval. A hospital buys a CT scanner, an MRI scanner or a laboratory analyser to use for well over a decade. In that time the operating system on the control console is replaced by several newer versions, and its developer stops issuing patches long before the hospital decides to replace the machine.

Several causes combine here, and none of them is negligence on the part of the IT department:

  • The device is approved as a whole. The hardware, operating system, application and configuration were verified together. A patch or an antivirus program that the hospital installs on its own can change how the machine behaves, and it usually breaches the warranty and service terms.
  • Manufacturer support has an end date. After that date the hospital receives an offer for a paid upgrade or a replacement, even though the machine still works correctly from a clinical point of view. Whether the purchase happens depends on the investment plan.
  • The machine works without breaks. A maintenance window in a unit that examines patients from morning to night means rescheduling examinations.
  • The device and the network have different owners. The IT department often has no administrator account on the console. The device is looked after by the clinical engineering department and the manufacturer’s service team.

An old system in a device is therefore a known risk that has to be managed: you describe it, you limit it and you plan its end.

Which networked devices need the most attention

Consoles, intermediate servers and workstations work in the network alongside the device itself. They have access to patient data, and they are often missing from the equipment register.

Device group Typical weakness Effect of a failure or compromise
Diagnostic imaging (CT, MRI, X-ray, ultrasound) with consoles Old console operating system, shared service password, open file shares Imaging unit out of service, leaked examinations, an entry point to the rest of the network
Laboratory analysers Intermediate computer also used for e-mail and web browsing Results held up, data retyped by hand
Patient monitors, central monitoring stations, ventilators Same network as office computers, wireless network with a single shared password Loss of the parameter view at the central station, disruption on the ward
Reporting workstations, disc burners, integration gateways Treated as ordinary computers although they have access to examinations Data leak, spread of malware

What the manufacturer is responsible for: medical device cybersecurity under the EU MDR

The manufacturer is responsible for building the device software in line with the current state of the art in security, and for telling the user under which network conditions the device may be operated. Both duties come from Regulation (EU) 2017/745 (MDR), which applies in every Member State. Annex I, Section 17.2 requires software to be developed and manufactured in accordance with the state of the art, taking into account the life cycle, risk management including information security, verification and validation. Section 17.4 obliges the manufacturer to set out minimum requirements for hardware, IT network characteristics and IT security measures, including protection against unauthorised access, that are necessary to run the software as intended. For in vitro diagnostic devices the corresponding act is Regulation (EU) 2017/746 (IVDR).

The regulation is supplemented by the MDCG 2019-16 Rev.1 guidance on cybersecurity for medical devices. It covers the stage before a device is placed on the market and the stage after it. On changes and patches it says they should be applied in line with the manufacturer’s guidance.

The Cyber Resilience Act (CRA, Regulation (EU) 2024/2847) does not apply to products covered by the MDR and the IVDR, so its requirements do not extend to equipment that is a medical device. A device manufacturer can, however, be subject to cybersecurity law in its own right. In Poland, the manufacture of medical devices and in vitro diagnostic devices is listed in Annex 2 to the KSC Act (important sectors), and a manufacturer that is at least a medium-sized enterprise has the status of an important entity.

A hospital can therefore expect three things from the manufacturer: a document with the network requirements and recommended security measures, a list of the connections the device needs, and information about the operating system version and the end-of-support date. No answer about a machine bought many years ago is also information. Such a device goes into the highest risk group and gets the tightest isolation in the network.

What is the hospital responsible for as the operator of the equipment?

The hospital is responsible for the environment the device works in: the network, the accounts, service access, configuration backups, monitoring, and the decision about what to do with a risk the manufacturer will not remove. This division holds in any country. What differs between countries is the legal basis.

Two EU instruments set the frame. NIS2 (Directive (EU) 2022/2555) is a directive, so the concrete duties of a hospital come from the national law that transposes it, and you have to check that law in each country where you operate. GDPR applies directly everywhere: the examinations and results stored on devices are health data, and Article 32 on the security of processing applies to them whether or not the provider is in scope of NIS2.

In Poland: what the KSC Act requires

For providers with the status of an essential or important entity, the duties come from Article 8(1) of the amended KSC Act (in Polish). Equipment connected to the network is part of the information system used to provide the service, so the following duties, among others, cover it:

  • asset management (Article 8(1)(2)(m)), which means knowing which devices work in the network;
  • security and continuity of the supply chain (point (e)), including the relationship with the manufacturer and the service company; the assessment of a supplier takes into account the vulnerabilities linked to it and the overall quality of its products (Article 8(2));
  • continuous monitoring (point (g)) and access control policies (point (n));
  • regular software updates in line with the manufacturer’s recommendations, taking into account an analysis of the update’s impact on the security of the service provided and the criticality of individual updates (Article 8(1)(5)(b), our translation).

The last point matters most for clinical equipment. The Act does not tell you to install every patch immediately. It tells you to follow the manufacturer’s recommendations and to analyse the impact first, which is compatible with the MDR rules. It does, however, require immediate action once a vulnerability is noticed (Article 8(1)(5)(d)). When a patch cannot be installed, you have to apply other safeguards, for example isolating the device in the network.

Entities that met the criteria on 3 April 2026 have until 3 April 2027 to implement the obligations. Not every hospital is covered, because size and legal form decide. We explain the classification rules in our article on which hospitals, clinics and labs are in scope of NIS2 in Poland, and the free Regulatory Compass survey gives a first answer.

A Polish provider outside the KSC Act is not released from looking after its devices. Besides Article 32 of GDPR, § 1(5) of the medical records regulation (Regulation of the Minister of Health of 6 April 2020, consolidated text Journal of Laws 2026, item 1391) requires systematic risk assessment, security measures adequate to the threats and attention to software updates in the systems that process medical records.

Who does what inside the hospital

Role Responsibility for clinical equipment
Head of the entity (the management board or managing director, or the director of a public hospital) Approves the rules, allocates funds, accepts in writing the risk that cannot be removed. In an entity covered by the KSC Act, the head remains responsible for meeting the obligations even after entrusting the tasks to another person (Article 8c).
Clinical engineering department Device register, contact with the manufacturer, inspections, requests for patches and for end-of-support information.
IT department, network administrator Segmentation, firewall rules, remote access, configuration backups, log collection.
Person responsible for information security Risk assessment, exception register, review of service access, report to the head of the entity.
Head of the imaging unit, laboratory or ward Downtime procedure, the clinical decision to stop using a device, reporting unusual device behaviour.

If nobody in the organisation combines the IT, clinical engineering and management perspectives, an external security lead working in the vCISO model can fill that role.

How do you inventory the equipment connected to the network?

The best way to start an inventory is to compare two lists that exist separately in most hospitals: the device register kept by clinical engineering and the list of addresses visible in the network. The difference between them shows the devices that one side does not know about.

  1. Take the equipment register and mark the items that have a network interface: wired, wireless or a modem.
  2. Collect from the switches and the address-assignment (DHCP) server the list of devices actually present in the network.
  3. Complete the picture with passive traffic observation, that is, analysis of a copy of the traffic taken from a switch. It shows who talks to whom without sending anything to the devices.
  4. Create a record for each device and assign two owners: a clinical owner (the head of the unit) and a technical owner (clinical engineering or IT).
  5. Assign a risk class on a three-level scale, considering the impact on the patient, the support status and the range of connections.
  6. Decide when the register is updated: with every new device, every service visit and every change in the unit’s network.

A device record contains at least: location, model and serial number, network address and segment, operating system version with its end-of-support date, the list of connections (system, port, direction), whether patient data is stored locally, local and service accounts, the method of remote servicing and the date of the last configuration backup.

Active network scanning can freeze an older device, so the equipment segment is not scanned the way an office network is. Vulnerability scanning in this part of the network needs agreement with clinical engineering, a gentle scan profile and a time slot outside examination hours.

A typical result of the first comparison in a CT unit: the register lists one machine and the network shows several addresses. They belong to the operator console, the reconstruction server, the reporting workstation, the contrast injector and a service router from the manufacturer that the IT department had not been told about.

How do you separate medical devices from the rest of the hospital network?

You separate clinical equipment by moving it into its own network segments and allowing only the connections that the manufacturer and the clinical process require, with all other traffic blocked by default. The goal is a network in which an infected computer in administration cannot see the CT console, and a compromised console cannot see anything except its image archive.

  • Segments by function. Diagnostic imaging, the laboratory, patient monitoring in the intensive care unit and the operating theatres, and infusion devices each get a separate segment. Service access gets a zone of its own.
  • Traffic between segments goes through a firewall. The rules take the form of a list of permitted connections. Everything that is not on the list is dropped and logged.
  • No internet access from equipment segments. The exception is documented connections to the manufacturer’s addresses, if remote diagnostics requires them.
  • No direct access from the office network to consoles. A doctor views examinations from the archive, not from the machine.
  • Control of wall sockets and the wireless network. A network socket in the unit is assigned to a specific segment, and monitors and pumps that use radio connectivity do not share a network with staff laptops and phones.

An example connection list for a CT scanner: the console sends images to the archive (PACS) using the DICOM standard, fetches the worklist from the imaging unit’s system and synchronises its clock with the hospital’s time server. The service team connects to the console only through a jump host in the service zone. E-mail, the web browser and file shares from other departments are blocked.

The rollout needs care, because one wrong rule stops the unit. A safe sequence is: several weeks of observing traffic without blocking, writing the rules from those observations and the manufacturer’s documentation, switching on blocking in an agreed window with the head of the unit present, and then testing the whole path with a trial examination, a transfer to the archive and a read on the reporting workstation. The largest cost is the time it takes to establish what really communicates with what. If your team does not have that time, the project can be handled as part of outsourced network administration.

Segmentation also limits the reach of an encryption attack. We develop that point in our article on how to keep admitting patients during a ransomware attack, and we describe the role of the perimeter firewall in our text on UTM systems.

Legacy systems: what do you do with a device that cannot be updated?

A device that cannot be updated may stay in use, provided you apply compensating controls and document the decision to accept the risk, with a review date. Switching off a clinically sound machine is rarely realistic, and doing nothing cannot be defended in front of an auditor.

The procedure always follows the same order. First comes a written question to the manufacturer about a patch or an upgrade path. The answer, including a refusal, goes into the device record. Next come the compensating controls from the table below. Where they concern the console itself, the service team carries them out or approves them. Finally you draw up an exception record.

Risk Control Who carries it out
Console operating system without patches Isolation in a segment with a minimal connection list, traffic filtering with signatures of known attacks at the entry to the segment, an exception record IT, clinical engineering, head of the entity
Shared or default service password Passwords changed in agreement with the manufacturer, named accounts on the access gateway Service team, IT
Permanent service tunnel or manufacturer’s modem On-demand access through the hospital’s gateway, multi-factor authentication, session recording IT, manufacturer
Removable media and service engineers’ laptops A station for checking media, rules written into the contract, a service port in a dedicated segment Clinical engineering, service team
Patient data left on the device’s disk Short local retention, data erased before repair outside the hospital and before disposal Clinical engineering, service team
No configuration backup A backup after every change, a restore attempt at each inspection Service team, IT

An exception record contains a description of the device, the reason it cannot be updated, the controls applied, an assessment of the remaining risk, the owner, the date of the next review and the signature of the head of the entity. The list of exception records also feeds the investment plan, because it shows which machines should be replaced first. In Poland, in entities covered by the KSC Act, planning adequate funds is one of the head’s tasks (Article 8d).

How do you give the manufacturer’s service team safe access to a device?

Remote servicing should enter the network only through a gateway the hospital controls, on a named account with multi-factor authentication, for the duration of a specific service ticket and with the session recorded. With permanent manufacturer tunnels, service routers plugged in beside the firewall and remote desktop tools installed on the console, the hospital does not know who is working on its machine or when.

The target flow of a service request looks like this:

  1. The service team reports the need for a connection, and clinical engineering confirms the time with the head of the unit.
  2. The IT department enables access for a named person, to a named device, for a set period.
  3. The service engineer logs in to the access gateway with a second factor, and from there to a jump host from which only that engineer’s device is visible.
  4. The session is recorded, and the staff in the unit know it is in progress.
  5. When the window ends, access expires automatically. The engineer hands over a report of the changes, and clinical engineering updates the device record.

Telemetry, meaning diagnostic data the device sends to the manufacturer, can be permitted as an outbound connection to specified addresses, once you have established in writing whether it includes patient data. For on-site servicing, the engineer’s laptop is connected to the device’s service port or to a dedicated segment, never to the office network.

If the service engineer has access to patient data, a legal basis is needed. When a service company processes data on behalf of the hospital, a processing agreement under Article 28 of GDPR comes into play. In Poland, Article 24(2)(2) of the Patient Rights Act also allows data from medical records to be processed by persons who perform activities related to maintaining the IT system and ensuring its security, on the basis of an authorisation from the data controller, and those persons are bound by secrecy. Which of the two to use for a specific device and contract should be confirmed with your data protection officer. We describe the rules for granting permissions in our article on who can access patient data.

Medical device cybersecurity day to day: monitoring and incident response

Clinical equipment has one property that makes oversight easier: its network traffic is repetitive. A CT scanner talks to the same systems every day, so every deviation deserves a check: a connection attempt from a console to the internet, address scanning from inside the segment, a service login outside a registered ticket, or a new device in one of the unit’s sockets.

The data comes from the logs of the firewall between segments, the switches, the service gateway and a passive sensor. The logs should go to a central SIEM system, and the alerts to a team that also responds at night, that is, an in-house or external 24/7 SOC. In Poland, for entities covered by the KSC Act, this is how the duty of continuous monitoring is met.

The second pillar is knowledge about vulnerabilities: manufacturers’ advisories and the warnings issued by the sectoral response team. In Poland that team is CSIRT CeZ, the sectoral CSIRT for health run by Centrum e-Zdrowia (the e-Health Centre, CeZ). According to CSIRT CeZ data presented on 1 October 2026, vulnerable services were the second most frequent category of events handled in the Polish health sector in 2025 (344 of 1,441). Since the beginning of 2026 the team’s experts have described 285 vulnerabilities with recommendations and detected 70 vulnerabilities in the infrastructure of health-sector entities (CeZ announcement, in Polish). The figures concern the whole infrastructure of healthcare providers, not clinical equipment alone.

When a device behaves suspiciously

The procedure has to keep two decisions apart. Cutting a device off from the network is a technical decision and usually a reversible one: many devices can work locally and send their examinations later. Stopping a device while it is in use with a patient is a clinical decision, and it belongs to the medical staff. The next steps are:

  • isolate the segment or the single socket at network level, without switching off the device’s power unless patient safety requires it;
  • move the unit to a downtime mode that it has rehearsed beforehand;
  • notify the manufacturer and clinical engineering, and preserve the logs of the firewall, the service gateway and the device itself;
  • assess whether the event is a serious incident. In Poland, an essential or important entity then submits an early warning within 24 hours of detection and an incident notification within 72 hours to CSIRT CeZ. If personal data has been breached, a separate 72-hour deadline for notifying the data protection authority under GDPR also runs. In Poland that authority is UODO.

We describe the reporting procedure step by step in our article on how a healthcare provider reports an incident to CSIRT CeZ.

What should you require when buying new equipment?

When buying equipment, require documented information about the device’s security from the supplier, together with commitments for the whole period of use, because both are hard to obtain once the contract is signed. The IT department and the person responsible for security should give their opinion on the purchase before the tender is announced, not on the day of installation.

  • A document with the device’s network requirements and security measures, handed over at the offer stage, together with a list of connections, ports and traffic directions.
  • The operating system version and the declared period for which security patches will be supplied, set against the planned period of use, together with the terms of an upgrade after that period.
  • A deadline for informing the hospital about a vulnerability and a deadline for delivering an approved patch.
  • Remote servicing on the hospital’s terms: access gateway, named accounts, a second login factor.
  • The ability to change default passwords and to forward logs to the hospital’s system.
  • A configuration backup, a restore procedure and a declared time for bringing the device back into operation.
  • Erasure of patient data when the device is repaired outside the hospital, returned or disposed of.

In public procurement, the way these requirements are written into the description of the subject of the contract and the evaluation criteria has to be agreed with your procurement department. This article does not settle questions of public procurement law. For more on contract clauses, see our article on what hospitals will require from medical software vendors, written from the vendor’s side of the table.

Where to start: a plan for this week and this quarter

You can start putting medical device cybersecurity in order without a budget, at the cost of a few hours of joint work by the IT department and clinical engineering.

This week:

  • compare the equipment register with the list of addresses in the network and count the differences;
  • find every permanent service connection: manufacturers’ routers, modems and tunnels;
  • block e-mail and web browsing on consoles that do not need them;
  • decide and write down who takes the decision to cut a device off from the network during an incident.

This quarter:

  • create device records for diagnostic imaging, the laboratory and the intensive care unit;
  • start the first segment in observation mode, and switch on blocking once the rules are written down;
  • replace permanent tunnels with a service gateway and named accounts;
  • ask manufacturers about support status and present the exception records to the head of the entity for signature;
  • rehearse two hours of work in one unit without the network, following the downtime procedure.

The most common mistakes in projects like this are starting with the purchase of a tool instead of the register, scanning equipment without agreeing it with the unit, rolling out segmentation in one day without an observation period, and leaving service access out of the project because “that is the manufacturer’s business”.

FAQ: frequently asked questions

Can a hospital install an operating system update on a medical device itself?

As a rule it should not do so without the manufacturer’s consent. The device was verified in a specific configuration, and the MDCG 2019-16 guidance says that changes and patches are applied in line with the manufacturer’s guidance. The hospital requests an approved patch in writing and, until it arrives, uses network isolation and other compensating controls.

Does a medical device with an old operating system have to be taken out of service?

The rules discussed in this article contain no such order. They require a risk assessment and measures adequate to the threats. The device can keep working if it is separated in the network and its service access is under control, and if the decision to accept the remaining risk has been documented and has a review date.

Does Poland’s KSC Act cover medical devices connected to the network?

Yes, if the provider is an essential or important entity. The obligations in Article 8(1), including asset management, the supply chain, updates and monitoring, apply to the information system used to provide the service, and networked equipment is part of that system.

Who is responsible for medical device cybersecurity: the manufacturer or the hospital?

The responsibility is shared. The manufacturer answers for the secure design of the software and for setting out the network and security requirements (MDR Annex I, Sections 17.2 and 17.4). The hospital answers for the network, the accounts, service access, monitoring and incident response.

How should remote service access to a CT scanner or an analyser be set up?

Access should go through a gateway controlled by the hospital, not through a permanent manufacturer tunnel. The service engineer uses a named account with multi-factor authentication and gets access only to the relevant device and only for the duration of the ticket. The session is recorded.

Does the Cyber Resilience Act (CRA) apply to medical devices?

No. Regulation (EU) 2024/2847 does not apply to products covered by the MDR and the IVDR. The requirements for medical device software come from Annex I to the MDR.

How we can help

Remote Admin works with networks of healthcare facilities. We help inventory the equipment visible in the network, roll out segmentation in a way that limits the risk of downtime in clinical units, put service access in order and bring the equipment segments under monitoring. Every engagement gets a custom quote. If medical device cybersecurity in your organisation needs an assessment of the current state, contact us about an audit.